Ease of use is at the heart of our innovations; we strive to ensure our products are intuitive to use, allowing the clinician to concentrate on the job in hand. The touch-screen technology of the E50 electrosurgery system takes the ease of use to a whole new level, minimising training requirements and facilitating fast set-up.
The E50 features leading edge technology in the form of Vessel Sealing and Argon Coagulation. Bipolar Vessel Sealing allows the sealing of vessels and tissue bundles up to 7mm in diameter. Argon Beam is a monopolar technology using high frequency electrical current to deliver ionized argon gas to create a wide area of coagulation with a limited penetration depth of 3mm.
The E50 delivers exceptional versatility through its 24 research based cutting and coagulation modes, plus optional Vessel Sealing and Argon Beam modes never before available in such an intuitive package.
Safety Features |
|
| Protection against electric shock: | |
| Class | I |
| Grade | CF |
| Low-frequency leakage currents | according to IEC 601-1-1 |
| High-frequency leakage currents | according to IEC 601-2-2 |
| Generator operation frequency | 357 kHz |
| Defibrillation impulse resistance | according to EN 60601-1 |
| NEM System | neutral electrode monitor |
| AUTOTEST | self-test of generator after switching the power on, service codes display |
| OVERLOAD | generator overload protection |
Monopolar Cut |
|
| Precise | 80W |
| Pure cut | 350W |
| Precise Cut | 80W |
| Blend 8 Levels |
180W |
| Cut with drying | 150W |
| Specialist cut | 350W |
| Argon cut | 350W |
| Papillotomy/Polypectomy 9 levels |
150W |
Monopolar Coagulation |
|
| Soft coagulation | 200W |
| Forced coagulation | 200W |
| Forced coagulation II |
80W |
| Spray coagulation | 80W |
| Argon coagulation | 80W |
| Argon pulse coagulation | 80W |
Bipolar Cutting |
|
| Bipolar cutting (4 different levels of haemostasis) |
100W |
| Bipolar specialist cut | 350W |
Bipolar Coagulation |
|
| Bipolar coagulation | 120W |
| AutoStart / AutoStop | automatic start and stop of bipolar coag. |
Argon Beam |
|
| Gas type | pure argon 5.0 (99.999%) or 4.8 (99.998%) |
| Inlet gas pressure | 0.25-0.4MPa |
| Gas outflow | 0.1-9.9 l/min, regulated 0.1 l/min in whole range |
| Pressure gauge | reducer (0.4MPa) on argon cylinder |
Vessel Sealing |
|
| Vessel sealing system | 250W |
| AutoStop | automatic stop of sealing process after achieving the optimal effect |
| Effect regulation | 8 levels of intensity regulation |
Weight & Dimensions |
|
| Length | 420mm |
| Width | 380mm |
| Height | 160mm |
| Weight | 12kg |
Power Supply |
|
| Supply voltage | 230V 50/60Hz |
Other Features |
|
| PowerStart system | Boosts the beginning of cutting process |
| Power regulation system | Automatic power regulation depends on needs |
| PP endoscopic programme | Includes endo detect |
| Programme memory | 104 different memory locations |
| MultiSwitch |
Remote programme selection by footswitch |
| Outputs |
2 monopolar, 2 bipolar |
Modular system upgrades allow users to meet future demands which may require the use of enhanced cutting and coagulation techniques by accessing integral argon beam and specialist vessel sealing modules.
Thermal sealing using vessel sealing forceps allows quick and effective sealing of large vessels and tissue bundles up to 7mm in diameter. Vessel sealing eliminates the need for traditional staples and sutures, by creating a permanent seal which can reduce both surgery and recovery times.
Argon plasma coagulation and argon enhanced cutting ensures fast, effective devitalisation of tissues which causes less blood loss and tissue damage.
![]() | E50 Electrosurgery System |