SES 2000 Autoclave
SES 2000 Specification

 

  Size Short
  Overall 46cm W x 46cm L x 36cm H
  Chamber 20cm dia. x 35cm L
  Tray Capacity 18cm x 28cm (2 off), 18cm x 14cm
(2 off), 1 of each supplied
  Tray Loading 1.5kg per tray
  Total Load 6kg per cycle
 
Weight
  Nett 27kg
  Shipping 32kg
 
Construction
   
  Chamber StainlessSteel
  Chamber Capacity 11 litres
  Reservoir Capacity 2 litres
  Door Stainless Steel
  Design Pressure 262kPa Gauge
 
Sterilizing
   
  Sterilising time @ 134°C 3 minutes
  Sterilising time @ 121°C 15 minutes
  Number of programmes 4
  Typical Overall Cycle Times 12 mins
  at 134°C 24 mins
  Drying Time Up to 17 minutes after sterilizing
 
Operating Pressure 137°C
232kPa (2.32 bar)
 
134°C
205kPa (2.05 bar)
 
124°C
125kPa (1.25 bar)
 
121°C
105kPa (1.05 bar)
 
Electrical
   
  Supply 230V ac 50/60Hz
  Loading 2kW @ 8.4A